BOWIE-DISK TEST SHEET

  • Axis Bowie-Dick test allows testing of steam sterilizers with minimum effort and highest level of quality.
  • Test pack helps to determination of sterilization malfunctions at an early stage.
  • The test sheet has a blue pattern with a lead-free chemical indicator.
  • Control of Steam sterilization process: 134°C – 3.5 min // 121°C-15 min

Once the cycle is over, the user must Analyse the results

INSTRUCTION FOR USE

1. Each day, before the first processed load, the Bowie-Dick test must be performed in an empty sterilizer to detect the presence of residual air in the chamber. Bowie-Dick Test is performed after a warm-up cycle at the beginning of every load.
2. Place Bowie-Dick Test horizontally on bottom shelf of sterilizer rack, over drain. Do not place directly on the drain or sterilizer floor.
3. Run the sterilizer at 134°C – 3.5 min // 121°C-15 min. 
4. Remove the test from the sterilizer and wait until it cools.
5. A uniform color change to black or its shades indicates “PASS”. Non- uniform color changes indicates “FAIL”.
6. “PASS” result indicates successful sterilization “FAIL” result indicates malfunction in sterilization process.

COMMON CAUSES OF ERRORS

  • Low pressure in the chamber
  • Steam supply changes
  • Leakages
  • Poor steam quality
  • Insufficient vacuum

The contents of each pack must be exposed to several critical parameters of the sterilization process.

    • Time
    • Temperature
  • If the ink has changed into another color (as specified), the items have been processed correctly.
  • Alternatively, any different colors (as specified) indicate a failure in the sterilization cycle.

The contents of each pack must be exposed to several critical parameters of the sterilization process.
-Time
-Temperature

Designed under Quality Management System standards ISO 11140-1 ISO 11140-3

T = 10-30 ºC, RH = 30-80 %, keep out of direct light.

50 Sheet.
Packing information: product description, instructions for use, regulation, storage conditions, manufacturer information and data on label.

  • Hot surface. Please be careful to avoid skin burns before opening the pack
  • Do not store the product near sterilizing agents.
  • Do not expose this product to EO, Dry Heat or any sterilization process other than Steam.
  • Do not use a Bowe- Dick Test Pack n gravity displacement steam sterilization cycles (where there’s no pre-vacuum phase to check) or any kind of sterilization processes other than vacuum-assisted Steam.
  • Do not reuse the sterilizer until the chemical indicator has changed homogeneously into the color indicated in the product.