BT40/6 is a biological indicator strip containing Geobacillus stearothermophilus spores (1 × 10⁶ CFU) used to monitor the efficacy of steam sterilization (autoclave) cycles. It is designed for routine control, validation, and qualification of moist heat sterilizers.
- Home
- Chemical Products
- BT40/6
BT40/6
- INSTRUTIONS OF USE
- AREAS OF APPLICATION
- FEATURES & BENEFTS
- COMPOSITION
- PRECAUTIONS OF USE
- PACKING TYPES
- PRODUCT CODE
INSTRUTIONS OF USE
- Place one BT40/6 strip inside a representative test pack or in the most challenging location of the autoclave load (hardest area for steam penetration).
- Run the sterilization cycle following your validated parameters.
- After the cycle, under aseptic conditions, transfer the strip into a sterile tube containing a suitable culture medium (e.g. Terragene’s Tryptic Soy Broth or Bionova medium).
- Incubate at 56–60 °C for up to 7 days, or as per your protocol.
- Observe for color change or turbidity indicating microbial growth:
- No growth (no color/turbidity) → Sterilization successful.
Growth (turbidity or color change) → Sterilization failed.
AREAS OF APPLICATION
- Monitoring and validation of steam sterilizers (autoclaves).
- Use in:
- Hospitals and health care facilities.
- Pharmaceutical and biotechnological industries.
- Laboratories and research centers.
- Medical device and instrument manufacturers.
FEATURES & BENEFTS
- Spore strips inoculated with Geobacillus stearothermophilus (1×10⁶ CFU).
- Excellent resistance and reproducibility according to ISO 11138-3.
- Ideal for routine control, qualification, and validation of autoclaves.
- Long shelf life with stable spore population.
- Compatible with various culture media.
- Reliable indication of sterilization efficiency.
COMPOSITION
- Paper carrier strip inoculated with Geobacillus stearothermophilus spores.
- Packaged in a protective glassine envelope to maintain sterility and stability.
PRECAUTIONS OF USE
- Store at 15–30 °C and dry conditions.
- Do not expose to sterilizing agents or moisture before use.
- Handle aseptically when transferring to growth media.
- Do not use damaged or expired strips.
- Dispose of after incubation according to biohazard waste procedures.
PACKING TYPES
- Packing: Box of 100 strips (in glassine envelopes).
PRODUCT CODE
- Product Code: BT40/6