IT26-C

02 Details

IT26-C is a chemical indicator integrator designed to monitor steam sterilization processes.
It is a Class 5 integrating indicator according to ISO 11140-1, meaning it reacts to all critical parameters of steam sterilization — time, temperature, and saturated steam — in a manner that correlates closely with the biological indicator performance.
It provides a precise visual confirmation that sterilization conditions were met inside the pack or load.

INSTRUTIONS OF USE

  1. Place one IT26-C indicator inside each pack, tray, or load before running the sterilization cycle.
  2. Process the load in the autoclave under your validated parameters (e.g., 121°C for 15 min or 134°C for 3.5 min).
  3. After the cycle, inspect the indicator window:
    • The blue bar must completely move into the “ACCEPT” area to confirm that sterilization conditions were achieved.
    • If the bar remains in the “REJECT” area, or incomplete → sterilization parameters not reached.

Keep the used indicator as documentation for sterilization records.

AREAS OF APPLICATION

  • Monitoring and control of steam sterilization cycles in:
    • Hospitals and clinical CSSDs (Central Sterile Services Departments)
    • Pharmaceutical and biotechnology facilities
    • Dental and veterinary sterilization units
    • Laboratories and industrial sterilization processes

FEATURES & BENEFTS

  • Class 5 Integrating Indicator — reacts to all critical parameters of steam sterilization.
  • Provides equivalent response to biological indicators with rapid visual readout.
  • Ensures accurate verification inside each pack, not just in the chamber.
  • Compatible with all standard steam sterilization cycles.
  • Clear, easy-to-read color movement between “Reject” and “Accept” zones.
  • Supports traceability and quality assurance programs.
  • Complies with ISO 11140-1 and EN 867-5 standards.

COMPOSITION

  • Indicator printed on special high-temperature-resistant card.
  • Contains chemical ink formulation that migrates based on sterilization parameters (time, temperature, and steam).
  • Non-toxic, latex-free materials.

PRECAUTIONS OF USE

  • Store in a cool, dry environment (15–30 °C) away from sterilants, heat, and humidity.
  • Do not use damaged or expired indicators.
  • For single use only — do not reuse.

Keep away from chemical vapors or sterilant gases before use.

PACKING TYPES

  • Packing: Box of 200 integrators (individually cut cards).

PRODUCT CODE

  • Product Code: IT26-C